CoaguChek Professional Monitoring Systems
Training Manual
35
CoaguChek XS Pro
Quality Control Testing
1) Remove plastic cap
and rubber bung
2) Ensure no liquid is in
the end of pipette and
carefully cut end off and
empty all contents into
glass vial, ensuring not to
touch the powder
3) Replace the black cap
and swirl on desktop
(do not shake), until
all the powder has
dissolved. Solution now
stable for 30 minutes
4) Touch ‘Control Test’
5) Insert test strip when
prompted and test strip
code chip (if prompted)
6) Check lot number on
side of QC bottle then
select from list, or
select ‘New’
7) If ‘New’ option
selected, insert
code chip when
prompted
8) Once the 180 Second
countdown starts, apply
the solution to the test
strip using the pipette
9) The sample will be
analysed and the
result and acceptable
range displayed on
the screen
Summary of Contents for CoaguChek XS Plus
Page 1: ...CoaguChek Professional Monitoring Systems Training Manual...
Page 4: ......
Page 5: ...CONTACT AND ORDER DETAILS...
Page 6: ......
Page 8: ......
Page 9: ...SYSTEM INTRODUCTION...
Page 10: ......
Page 16: ......
Page 17: ...INTERFERENCES...
Page 18: ......
Page 22: ......
Page 23: ...TRAINING...
Page 24: ......
Page 28: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 37: ...CLEANING...
Page 38: ......
Page 44: ......
Page 45: ...TROUBLESHOOTING...
Page 46: ......
Page 49: ...APPENDIX 1...
Page 50: ......
Page 65: ...APPENDIX 2...
Page 66: ......
Page 67: ...Accuracy and precision in oral anticoagulation monitoring I know my value...
Page 77: ......