Cleaning/disinfecting the test strip guide
Apply the solutions for a contact time of > 1 minute (refer to the corresponding product labelling).
Using lint-free cotton swabs/buds.
1. Remove the test strip guide cover to clean it. (Use your thumbnail to open the cover of the test
strip guide by pressing its front edge upward.) Move the cover safely away from the meter.
Then rinse the cover with warm water or wipe it clean using the solutions recommended above.
Let the test strip guide cover dry
for at least 10 minutes
before re-attaching it.
Then rinse the cover with warm water or wipe it clean using the solutions recommended
above. Let the test strip guide cover dry for at least 10 minutes before re-attaching it.
2. Hold the meter upright with the test strip guide facing down.
•
Clean the easily accessible white areas with a moistened cotton swab/bud.
•
Ensure the swab/bud is only damp, not wet. Wipe away residual moisture and fluids.
NOTICE
Damage to the instrument
Ensure that no liquid enters the meter. If moisture enters the meter, it may
cause malfunction of the instrument.
Do not insert any objects in the test strip guide. Doing so may damage the
electrical contacts behind the test strips guide.
On completion of cleaning:
3. With the cover off, let the test strip guide dry for at least 10 minutes before re-attaching the
test strip guide cover and testing again.
•
Visually verify that no residual moisture is seen anywhere on the test strip guide and
cover at the completion of cleaning and disinfecting.
•
Ensure that the test strip guide and cover are completely dry before assembling them.
2. Hold the meter upright with the test strip guide facing down.
• Clean the easily accessible white areas with a moistened cotton swab/bud.
• Ensure the swab/bud is only damp, not wet. Wipe away residual moisture and fluids.
Then rinse the cover with warm water or wipe it clean using the solutions recommended
above. Let the test strip guide cover dry for at least 10 minutes before re-attaching it.
2. Hold the meter upright with the test strip guide facing down.
•
Clean the easily accessible white areas with a moistened cotton swab/bud.
•
Ensure the swab/bud is only damp, not wet. Wipe away residual moisture and fluids.
NOTICE
Damage to the instrument
Ensure that no liquid enters the meter. If moisture enters the meter, it may
cause malfunction of the instrument.
Do not insert any objects in the test strip guide. Doing so may damage the
electrical contacts behind the test strips guide.
On completion of cleaning:
3. With the cover off, let the test strip guide dry for at least 10 minutes before re-attaching the
test strip guide cover and testing again.
•
Visually verify that no residual moisture is seen anywhere on the test strip guide and
cover at the completion of cleaning and disinfecting.
•
Ensure that the test strip guide and cover are completely dry before assembling them.
NOTICE Damage to the instrument
Ensure that no liquid enters the meter. If moisture enters the meter, it may cause
malfunction of the instrument.
Do not insert any objects in the test strip guide. Doing so may damage the electrical
contacts behind the test strips guide.
Summary of Contents for CoaguChek XS Plus
Page 1: ...CoaguChek Professional Monitoring Systems Training Manual...
Page 4: ......
Page 5: ...CONTACT AND ORDER DETAILS...
Page 6: ......
Page 8: ......
Page 9: ...SYSTEM INTRODUCTION...
Page 10: ......
Page 16: ......
Page 17: ...INTERFERENCES...
Page 18: ......
Page 22: ......
Page 23: ...TRAINING...
Page 24: ......
Page 28: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 37: ...CLEANING...
Page 38: ......
Page 44: ......
Page 45: ...TROUBLESHOOTING...
Page 46: ......
Page 49: ...APPENDIX 1...
Page 50: ......
Page 65: ...APPENDIX 2...
Page 66: ......
Page 67: ...Accuracy and precision in oral anticoagulation monitoring I know my value...
Page 77: ......