CoaguChek Professional Monitoring Systems
Training Manual
27
CoaguChek XS Plus
Patient Testing
1) Touch
‘Patient Test’
2) If prompted enter the
Patient ID via touch
screen then touch
3) Remove test strip
from pot and insert into
meter when prompted
4) If prompted
insert
code chip
5) The meter finds the code
chip information and warms
the strip then displays a 180
second countdown
6) Lance the
patients’
finger
7) Apply the drop of blood
to the test strip using
top or side dosing,
within 15 seconds
of lancing finger
8) Wait for the meter to
beep and start to analyse
the sample then remove
the finger. The meter will
automatically test the
onboard QC
9) The result is
displayed.
Comments can
be added
by touching