3-5.Using the Input Terminals
• All functions except current use the V/OHMS and COM input terminals.
• The two current input termin for 500mA and 10A current measurement.
10A:
Input for 0A to 10.00A current (20A overload for 30 seconds on, 10 minutes off).
μAmA:
Input for 0A to 500mA current measurements.
COM:
Return terminal for all measurements.
VΩCAP
Hz Temp:
Input for Voltage, Continuity, Resistance, Diode Test, Conductance,
Capacitance.
4.Measure Function
4-1.AC Voltage Measurement
WARNING:
Risk of Electrocution. The probe tips may not be long enough to contact the live
parts inside some 240V outlets for appliances because the contacts are recessed deep in the
outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it.
Make sure the probe tips are touching the metal contacts inside the outlet before assuming
that no voltage is present.
CAUTION:
Do not measure AC voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the
VAC
Position.
2. Insert the black test lead into the
COM
Input Jack and the red test lead into the
Positive
Input Jacks.
3. Position the black test probe tip and red test probe tip respectively in the spots of the
circuit to be measured.
4. Read the AC voltage in the display.
OFF
CAP
mA
AUT
O POWER OFF
10A
mA
A
10A
V
10A
800mA
FUSED
COM
CAT IV 600V
CAT III 1000V
True RMS Autoranging Digital Multimeter
30 sec MAX
every 15 min
IP67
WATERPROOF
V
V CAP
Hz
%
Temp
µA
AC+DC
Hz
%
Temp
°C °F
LPF
mA%
4~20
mV
Bluetooth
RANGE
HOLD
REL
F1
F2
F4
F3
5. 0000
Auto range
06:32
GRAPH
01
23
4
5V
MODE
SAVE
MAX
V
4-2.DC Voltage Measurement
CAUTION:
Do not measure DC voltages if a motor on the circuit is being switched ON or
OFF. Largevoltage surges may occur that can damage the meter.
1. Set the function switch to the
VDC AC+DC
Position.
2. Insert the black test lead into the
COM
Input Jack and the red test lead into the
Positive
Input Jacks.
3. Position the black test probe tip and red test probe tip respectively in the spots of the
circuit to be measured.
4. Read the DC voltage in the display.
OFF
CAP
mA
AUT
O POWER OFF
10A
mA
A
10A
V
10A
800mA
FUSED
COM
CAT IV 600V
CAT III 1000V
True RMS Autoranging Digital Multimeter
30 sec MAX
every 15 min
IP67
WATERPROOF
V
V CAP
Hz
%
Temp
µA
AC+DC
Hz
%
Temp
°C °F
LPF
mA
%
4~20
mV
Bluetooth
RANGE
HOLD
REL
F1
F2
F4
F3
5. 0000
Auto range
06:32
GRAPH
-5
5V
MODE
SAVE
MAX
V
-4
-3
-2
-1
01
23
4