Specifications
General specifications
Parameter
Value / Range
Power supply
100 - 240 VAC (50/60 Hz)
Protection class
II
Power consumption
15 VA (in operation)
Load capacity of the relay
outputs
2.5 A per relay
IP protection class
without front cover
IP43
with optional front cover
IP54
Storage temperature
0 °C - 45 °C
Ambient temperature
10 °C - 45 °C
Measured water temperature
5 °C - 40 °C
Humidity
20 - 90 % RF (without ice or condensation)
Inlet water pressure
min: 0.5 bar - max: 5 bar / recommended 1 - 2 bar
Inlet water in general
clear, colourless, free of solids, without gas bubbles
Water quality requirements for
the measurement of water
hardness
pH:
4 - 10
Iron:
< 3 ppm
Copper:
< 0.2 ppm
Aluminium:
< 0.1 ppm
Manganese:
< 0.2 ppm
Acid capacity:
KS 4.3 < 5 mmol/l
Note on oxidising agents:
Oxidising agents such as calcium hypochlorite, chlorine, chlorine dioxide, sodium hypochlorite or
ozone above the limits permitted in "TrinkwV 2012" attack the dye contained in the reagent and
interfere with the measurement. This means that an exact determination of the water hardness is
no longer guaranteed. An activated carbon filter connected upstream of the analyser can remove
these oxidants from the sample water and thus enable the correct determination of the water
hardness.
The capacity of an activated carbon filter is consumed during operation. Therefore, the activated
carbon filter must be replaced at regular intervals. The effectiveness of the activated carbon filter
can be checked with the help of Caldur® test sets.
Signal input and outputs
Parameter
Value / Range
Relay outputs
2 Relay
2.5 A / 250 VAC
2.5 A / 100 VDC
as potential-free outputs NC/NO
The relays provide the following functions:
•
Limit value alarm (relay output 1)
•
Device fault or reagent message BoB (relay output 2)
Input contact
(INPUT, start or stop
analysis interval)
only with LIMESPlus
Galvanically isolated input contact for connection of a potential-
free contact
•
Start analyses (external analysis interval)
•
Flow monitor (analysis interval interruption during system
standstill)