- 33 -
7.
Alarm Function
7-1. Gas alarm activation
When the concentration of detected gas reaches or exceeds the alarm setpoint values, a "gas alarm" is
triggered in the detector. The alarm lamp blinks, the buzzer sounds, and the bar meter display indicates an
alarm condition. (Auto-reset operation)
Blinking alarm lamp and buzzer sound operation for gas alarms have different intervals depending on the
detected concentration.
7-2. Fault alarm activation
"Fault alarm" is triggered using a buzzer sound and blinking alarm lamp when an abnormality is detected in
the detector. (Self-latching)
When an alarm is triggered, one of the following fault details is displayed on the LCD.
- System
abnormalities
: FAIL SYSTEM
- Battery voltage
low
: FAIL BATTERY
- Calibration
abnormalities
: FAIL AIR CAL
- Clock
abnormalities
: FAIL CLOCK
- Low flow rate
: FAIL LOW FLOW
- Pump
abnormalities
: FAIL PUMP
- Sensor
abnormalities
: FAIL SENSOR
Alarm lamp
Repeatedly blinks at about one-second intervals.
Buzzer
Repeatedly sounds intermittent beeps at about
one-second intervals.
Blip-blip, blip-blip
LCD display
Display example of low flow rate (LOW FLOW)
If a fault alarm is triggered, refer to trouble shooting and take appropriate action.
If the unit has problems and is repeatedly malfunctioning, contact RKI immediately.
NOTE
•
The low flow rate alarm (FAIL LOW FLOW), calibration abnormalities (AIL AIR CL) and clock
abnormalities (FAIL CLOCK) can be reset by pressing the MODE button.
•
For information on malfunctions (error messages), see "Troubleshooting" on page 41.