Setting Basic Operational Parameters in the Config Menu • 44
Calibration Parameters
There are five calibration parameters. Table 13 below describes the parameters and
shows the factory settings. If you wish to use the factory settings, then you do not
need to make any parameter adjustments. If you wish to confirm or change the
parameter settings, follow the instructions below.
Pre-Sample Flush Time (AIR Flush)
The pre-sample flush time is the length of time that the SDM-6000 will draw air
through the AIR fitting before performing an air adjust operation.
Calibration Gas Sample Time (GAS)
The calibration gas sample time is the length of time that the SDM-6000 will draw
calibration gas through each of the gas fittings.
Post-Sample Purge Time (AIR Purge)
The post-sample purge time is the length of time that the SDM-6000 will draw air
through the AIR fitting to purge calibration gas from the system.
Alarm Check
If Alarm Check is selected, a vibrator, LED, and buzzer test is done on the GX-6000
at the end of the bump test. The vibrator will vibrate, the LEDs will turn on for a few
seconds, the buzzer will sound, and the SDM-6000 will determine if these actions
were completed successfully.
Auto Print
The printer must be connected for the Auto Print parameter to have any function. If
Auto Print is selected, the SDM-6000 will automatically print calibration results at the
end of a calibration. If Auto Print is not selected, the SDM-6000 will not automatically
Table 13: Calibration Parameter
Parameter
Display Tag
Available Choices
Factory
Setting
Pre-Sample Flush
Time
AIR Flush
15 - 120 seconds in
1 second increments
15 seconds
Calibration Gas
Sample Time
GAS
20 - 120 seconds in
1 second increments
60 seconds
Post-Sample
Purge Time
AIR Purge
5 - 120 seconds in
1 second increments
15 seconds
Alarm Check
Alarm
Check
• On
• Off
On
Auto Print
Auto Print
• On
• Off
Off