66
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
KEY.TONE
item in
User Mode.
5. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Display Mode Items (DISP.SET)
OFF
:
USER ID
,
STN ID
, and
BUZZ.VOL
screens do not appear in Display Mode.
ON
(factory setting):
BUZZ.VOL
screen appears in Display Mode.
USER ID
and
STN ID
screens appear in Display Mode if
ID DISP
in Maintenance Mode is also set to
ON
(factory
setting is
OFF
).
1. While in User Mode, press AIR to scroll to
DISP.SET
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
DISP.SET
item in User
Mode.
5. See “Entering Measuring Mode (START)” on page 69 to enter Measuring Mode.
Zero Suppression (ZERO.SUP)
This item only appears if
ZSUP.DSP
is set to
ON
in Maintenance Mode (factory setting is
OFF
).
The
ZERO.SUP
setting is not intended for field adjustment. The default setting is
ON
. The
suppression value is 0.5% volume.
Zero Follower (ZERO.FLW)
This item only appears if
ZFLW.DSP
is set to
ON
in Maintenance Mode (factory setting is
OFF
).
The OX-04 does not support zero follower functionality. This menu item has no effect on
instrument operation.
DISP .SET