64
•
Chapter 5: User Mode and Calibration
OX-04 Operator’s Manual
BEEP.SEL
BEEP.SEL
defines what kind of confirmation or non-compliance indicator you want to occur in
Measuring Mode. The available choices are:
OFF
(factory setting): The OX-04 does not provide a confirmation alert or non-compliance
indicator.
LED
: The OX-04’s LED double flashes as often as defined by the
BEEP.INT
parameter to verify
that the instrument is operating.
BUZZER
: The OX-04’s buzzer double beeps as often as defined by the
BEEP.INT
parameter to
verify that the instrument is operating.
LED+BUZ
: The OX-04’s LED double flashes and the buzzer double beeps as often as defined by
the
BEEP.INT
parameter to verify that the instrument is operating.
BMP/CAL
: If a bump test or a calibration is due and if
BP.EXPD
or
CAL.EXPD
is set to
CONFIRM
(factory setting) or
NONE
, the OX-04’s LED double flashes as often as defined by
the
BEEP.INT
parameter to indicate a non-compliance. Once a bump test or calibration
(depending on which is due) is done, the LED stop flashing.
1. While in the
BEEP
menu, press AIR to scroll to
BEEP.SEL
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BEEP.SEL
item in the
BEEP
menu.
5. See “ESCAPE” on page 65 to return to User Mode.
BEEP.INT
The
BEEP.INT
parameter defines how often the confirmation alert or non-compliance indicator
selected in
BEEP.SEL
occurs. This setting only applies if the
BEEP.SEL
parameter is set to
something other than
OFF
(factory setting). The available choices are
0.5
minutes and
1
-
99
minutes in 1 minute increments. The factory setting is
5
minutes.
1. While in the
BEEP
menu, press AIR to scroll to
BEEP.INT
.
2. Press and release POWER MODE. The current setting flashes.
BEEP .SEL
BEEP .INT