OX-04 Operator’s Manual
Chapter 5: User Mode and Calibration
•
59
BP.RMDR
ON
: The OX-04 gives an indicator at start up if it is due for bump testing. The type of indicator
depends on the
BP.EXPD
setting (see page 60). If the instrument is not due for bump testing, a
check mark appears in the lower left corner of the LCD.
OFF
(factory setting): The OX-04 does not give an indicator at start up if it is due for bump
testing.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.RMDR
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.RMDR
item in the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.
BP.INT
This setting defines the amount of time between bump tests and can be set in 1 day increments.
The minimum setting is
0
days and the maximum setting is
30
days (factory setting).
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.INT
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.INT
item in the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 60 to return to User Mode.