86
•
Appendix A: Maintenance Mode
OX-04 Operator’s Manual
Turning the ID Display Function On/Off (ID DISP)
ON
: The User ID and Station ID screens appear in the startup sequence. If
DISP.SET
in User
Mode is also set to
ON
, then the IDs can be changed in Display Mode.
OFF
(factory setting): The User ID and Station ID screens do not appear in the startup sequence
and the IDs cannot be changed in Display Mode.
1. While in Maintenance Mode, press AIR to scroll to
ID DISP
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
ID DISP
item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.
Turning the Zero Suppression On/Off (ZERO.SUP)
The
ZERO.SUP
setting is not intended for field adjustment. The default setting is
OFF
. The
suppression value is 0.5% volume.
Turning the Zero Follower On/Off (ZERO.FLW)
The
ZERO.FLW
setting is not intended for field adjustment. The default setting is
OFF
.
User Mode Zero Suppression (ZSUP.DSP)
ON
: Zero suppression item appears in User Mode.
OFF
(factory setting): Zero suppression item does not appear in User Mode. The zero suppression
item is always available in Maintenance Mode.
It is not normally necessary to have the zero suppression item appear in User Mode. Contact RKI
Instruments before turning this setting on.
ID DISP