GX-2009 Operator’s Manual
Calibration Mode • 45
Program or the User Setup Program. To exit this screen and return to the main menu, press
and release the AIR button.
3. Use the sample tubing to connect the calibration adapter plate to the regulator. Attach the
tubing to the adapter plate on the inlet side as shown below in Figure 13.
Figure 13: Calibration Kit Assembly
4. Push the adapter plate onto the GX-2009’s sensor face. Make sure the adapter plate is
oriented as shown in Figure 14 below with the sensor names on the adapter plate matching
up with the sensor names on the instrument.
Figure 14: Installing the Adapter Plate