55
4
Push the PCB retaining tabs back,
grasp the cable socket, and lift the
sensor board out of the housing. Set
the sensor board aside.
5
Remove the top half of the main
unit’s housing to access the wiring at
the other end of the cable.
6
Remove the old cable’s wiring leads
from the sensor terminal strip.
For each terminal strip position:
-
Push the push button straight
downward using the compatible
screwdriver or equivalent to open
the spring.
-
Remove the wire.
-
Release the screwdriver.
7
Pull the old cable out of the housing.