Acting
※
Only gas alarm action is a lock-in system.
Response to gas alarm (In case of responding to leaked gas)
When a gas alarm is triggered, take actions in accordance with your management rules of gas alarm.
Normally, take the following actions.
- Check the reading of the monitor.
- If the gas alarm display continues to be displayed, close the main valve of the gas, and then
check that the gas concentration reading is back to normal.
NOTE
•
If a gas leak is momentary, the reading may already be back to normal when checking it. In addition,
when the alarm is triggered by noise or other incidental conditions other than a gas, the reading may
have already returned to normal.