62
8
8.
Storage and Disposal
8-1. Procedures to store the monitor or leave it for a
long time
The monitor must be stored under the following environmental conditions.
• In a dark place under the normal temperature and humidity away from direct sunlight
• In a place where gases, solvents or vapors, etc. are not present
Store the monitor in a shipping carton in which it was delivered, if possible.
Store the monitor away from dust, etc. if the shipping carton is not available.
8-2. Procedures to relocate the monitor or use it again
When the monitor is relocated, refer to "How to Install" for relocation sites and wiring work.
The unpowered time must be minimized when the monitor is relocated.
When the monitor is used again after long-term storage, perform a calibration.
8-3. Disposal of products
When the monitor is disposed of, it must be treated properly as an industrial waste in accordance with the
local regulations, etc.
Never fail to return the used sensor to RKI.
If liquid is leaked from the sensor, do not touch the liquid. The sensor must be put into a plastic bag to
prevent the liquid from leaking to the outside. If any liquid leaked from the sensor is found in the monitor
unit, turn off the power and contact RKI immediately.
CAUTION
•
If the monitor operating on dry batteries is not used for a long time, store it after removing the
batteries. Battery leaks may result in fire, injury, etc.
CAUTION
•
Contact RKI for information on readjustment including calibration.
WARNING
•
Do not disassemble the electrochemical type sensor because it contains electrolyte.
Electrolyte may cause severe skin burns if it contacts skin, while it may cause blindness if its
contacts eyes. If electrolyte is adhered on your clothes, the material will discolor or deteriorate.
If contact occurs, rinse the area immediately with a large quantity of water.
•
Dispose of dry batteries in accordance with procedure specified by the local authority.