36 • Chapter 6: Advanced Configuration Menu
AirLink 9850 Gas Monitor Operator’s Manual
Setting the Network ID
1. Press
ADD
or
SUB
to change the network ID.
NOTE:
All monitors and sensor assemblies must have the same Network Channel in order to
communicate.
2. Press
MENU
.
Adjusting the Primary/Secondary Status
1. Press
ADD
or
SUB
to switch the monitor to “Primary” or “Secondary”.
NOTE:
There can only be one Primary monitor on a network. All other monitors must be set up as
Secondary monitors.
If an AirLink 9850 is set as a Secondary monitor when there is no Primary monitor, the
AirLink 9850 will go into a Fault 15 status.
2. Press
MENU
to return to Normal Operating Mode.
Exiting the Advanced Configuration Menu
1. Press
RESET/ESC
at any time to exit the Advanced Configuration Menu.
2. Close the enclosure box.
3. Clamp down the enclosure latches.
NOTE:
The AirLink 9850 automatically saves any changes and returns to Normal Operating Mode
15 minutes after the last button press.