79
Store Operation
Auto (only when the VX-55EX is installed)
The processing result which is obtained using the selected vibration level and
specifi ed interval by the store interval setting will be recorded continuously.
When the following one of conditions occurred, the store is stopped and data
is saved.
• When the total measurement time reached the set value.
• When
the
L
eq
store reached 999,999 sets.
• When the capacity of the SD memory card became insuffi cient.
L
V
store
The vibration level and vibration acceleration level of up to 1,000 hours
can be stored continuously and automatically.
This function becomes available when an SD memory card is inserted.
This is useful when recording the vibration level waveform.
The store interval can be selected from 100 ms (milliseconds), 1 s (second).
L
eq
store
All processed data except vibration level and vibration acceleration level
are considered as a single data set, and up to 999,999 data sets can be
stored continuously and automatically.
This function becomes available when an SD memory card is inserted.
This is suitable for a measurement with a specifi ed long period of time.
The processing interval can be selected from 500 s (seconds), 10 s, 1 min
(minute), 5 min, 10 min, 30 min, 1 h (hour), 8 h, 24 h and manually selected
time (user setting; up to 24 hours).
Summary of Contents for VM-55
Page 2: ......
Page 14: ...2 Outline Block diagram...
Page 45: ...33 Preparations System screen Clock settings screen...
Page 47: ...35 Preparations System screen Backlight LCD settings screen...
Page 49: ...37 Preparations System screen Execution confirmation screen...
Page 51: ...39 Preparations Display I O screen Comparator screen...
Page 203: ......