168
Serial Interface
Manual Store
Manual store
Storing the calculated value in manual store
Setting command
Manual
Store, p1
Parameter
p1= “Start”
(Execute store)
There is no request command
Measurement Start Time
Measurement (operation) start time
Request measurement (operation) start time
Request command Measurement
Start
Time?
Response data
d1/d2/d3
d4:d5:d6
Returned value
d1= 2015 or after
(year)
d2= 1 to 12
(month)
d3= 1 to 31
(date)
d4= 0 to 23
(hour)
d5= 0 to 59
(minute)
d6= 0 to 59
(second)
There is no setting command
Measurement Stop Time
Measurement (operation) stop time
Request measurement (operation) stop time
Request command Measurement
Stop
Time?
Response data
d1/d2/d3
d4:d5:d6
Returned value
d1= 2015 or after
(year)
d2= 1 to 12
(month)
d3= 1 to 31
(date)
d4= 0 to 23
(hour)
d5= 0 to 59
(minute)
d6= 0 to 59
(second)
There is no setting command