176
Serial Interface
Command example
The example of a setting by a command is shown.
Using a request command after a setting is recommended.
Basic setting
Setting the display channel to “XYZ”
Display
Channel
Setting, XYZ
Setting the display characteristic type to “
L
v
”
Display
Lva
Setting, Lv
Setting the time scale of time-level display to “20s”
Time
Level
Time
Scale, 20s
Setting the level range upper limit of X channel, Y channel and Z channel to
“120 dB”
Level
Range
X, 120
Level
Range
Y, 120
Level
Range
Z, 120
When operating auto store
Setting the Store Mode to “Auto”
Store
Mode, Auto
Setting the Store Name to “0100”
Store
Name, 100
Setting the Total Measurement Time to “10min”
Measurement
Time
Preset
Auto, 10m
Setting the
L
v
Store Interval to “100ms”
Lv
Store
Interval, 100ms
Setting the
L
eq
Calculation Interval to “1min”
Leq
Calculation
Interval
Preset, 1m
Measurement start / stop (store)
Measure, Start
*
No confi rmation about overwriting.
Measure, Stop
Measurement data acquisition
DOD?