25
Measurement
Measurement
The KS-19F does not have measurement controls or a display for measurement
results. It is designed to perform measurement under control of a controller
KE-40B1 connected to it. For details, please refer to the documentation of
the controller KE-40B1.
The KS-19F is not equipped with a means for adjusting the fl ow rate of the
sample fl uid. Refer to the documentation of the fl ow rate adjustment device
in the measurement system and adjust the fl ow rate accordingly.
When the KS-19F is connected to the controller KE-40B1, particle counts
of the KE-40B1 is converted to number concentration with reference to an
effective sampling fl ow rate of the KS-19F.
Error message
When there is a problem with the KS-19F, the LIQUID LEAK indicator, CELL
indicator, and LASER indicator are used to indicate the condition.
The following explanation explains operation and error messages when the
problem has occurred.
For information on countermeasures, please refer to the section “Trouble-
shooting” on page 37.
Status of indicators on KS-19F
Status: KS-19F
status
Operation:
KS-19F operation and countermeasures
Output:
KE-40B1 display and serial interface output
Shows the *** part of the measurement error information
message having the format “E / ***<
EOL
>”.
When LIQUID LEAK indicator is lit in red
Status:
Leak has occurred inside cabinet.
Operation:
If measurement is in progress, it is stopped immediately
and measurement data are discarded.
Output: LIQUID_LEAK