17
6-4.Storage or treatment when not in use for a long period
Oxygen sensor is a kind of cell and will continue to consume its life even though it is not
used. When detector head is not used for a long period, take out the sensor from detector
head and store it at following environment to make lower the consumption of sensor.
・Environment of low temperature (5 - 10℃)
・Environment of low oxygen content
(Put a sensor into the container and seal off the container.)
6-5.List of recommendable spare parts for
regular replacement
No.
Name of parts
Inspection frequency
Replacement interval
Q’ty/unit
1
O-ring(sensor side in
the holder)
1 year
3-6 years
1
2
Packing(Housing side
in the holder)
1 year
3-6 years
1
* NOTE
・The replacement interval will change depending on operating condition and it does not
mean the warranty period.
Replacement interval changes according to the result of regular maintenance.