11
(2) In case of pressurized sample gas
(3) In case of aspirator suction
It is also possible to apply air aspirator unit to sample the gas by utilizing
instrumentation air. In this method, pressure to the sensor must be kept as
small as possible and make it to -20mmHg at a maximum.
As the load to oxygen sensor is high on this method, aspirator suction method
shall be used only when the pump cannot be used.
If the sample gas pressure would fluctuate, use the constant flow rate control
valve to keep the constant flow rate.
Exhaust the sampled gas to open air in principle. When make piping to exhaust
side necessarily, choose the piping diameter to keep the pressure loss due to
the length of piping within 10mmHg.
Dust filter
Detection
point
GD-F4A-SC-A
Gas OUT
1.0±0.3 L/min
Flow control valve
Flowmeter
CAUTION
!
Flowmeter
Detection
point
Instrumentation air
Air consumption
:1-3Nℓ/min
pressure:1-35kg/cm2G
Dust filter
Flow control
valve
Aspirator
GD-F4A-SC-A