5.HOW TO OPERATE
5-1.Preparation for start up
Before making power on, take care of the following. If do not keep this, there is the danger
of electrical shock and damage of instrument.
5-2.How to start the detector head
① This detector head does not equip a power switch. Analog outputs of oxygen sensor
are converted to 4-20mA and transmitted to the indicator/alarm unit directory
through the connected cable. As far as the cable has been connected correctly, this
system starts operation by putting on the power switch in indicator/alarm unit.
② Let the gas flow be sampled to this unit by turning on the pump switch etc.
③ Adjust the flowrate to lead sampled gas to 1.0L±0.3mL/min.
④ Be sure to perform gas calibration at the time of new installation. Correct reading
may not be obtained before gas calibration.
5-3.How to exit
① Stop the sample gas supply to this unit by turning off the pump switch, etc.
② This system stops operation by putting off the power in indicator/alarm unit, since this
unit has no power switch.
・ Make protective grounding.
・ Check that the wiring with outer unit is made correctly.
・ Check that the supply voltage is within the rating.
・ Check that the rated fuse is used.
WARNING
・At the new installation, adjust the reading in indicator/alarm unit according to
“
6-2.
Calibration
”
.
CAUTION
!