Return Format
The query returns the voltage of the specified channel, for example, 7.500.
Examples
:VOLT 7.5 /*Sets the voltage to 7.5 V for the current channel.*/
:VOLT? /*Queries the voltage for the current channel. The query
returns 7.500.*/
4.9.8
[:SOURce[<n>]]:VOLTage[:LEVel]
[:IMMediate]:STEP[:INCRement]
Syntax
[:SOURce
[<
n>]
]:VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
{<
voltage>|
<
def>}
[:SOURce
[<
n>]
]:VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]?
[<
def>]
Description
Sets or queries the step size of voltage for the specified channel.
Parameter
Name
Type
Range
Default
<n>
Integer
{1|2|3}
-
<voltage>
Real
0 to the maximum voltage value
of the specified channel
Please refer to
<def>
Discrete
{DEFault}
-
Remarks
•
If [:SOURce<n>] is omitted, the command sets the corresponding parameter of
the channel currently selected.
•
<voltage> is the step size specified. DEFault is the default value. The default
values of <voltage> are shown in the table below:
Channel
Default
CH1
0.001 V
CH2
0.001 V
CH3
0.001 V
Command System
64
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.