![Rigol DP2000 Series Programming Manual Download Page 64](http://html1.mh-extra.com/html/rigol/dp2000-series/dp2000-series_programming-manual_1461089064.webp)
Examples
:OUTPut:PAIR SERies /*Connects CH1 and CH2 in series (internal).*/
:OUTPut:PAIR? /*Queries the connection mode. The query returns
SERIES.*/
4.8.15
:OUTPut[:STATe]
Syntax
:OUTPut[:STATe]
[<
source>,]<bool>
:OUTPut[:STATe]?
[<
source>]
Description
Sets or queries the on/off state for the specified channel.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CH1|CH2|CH3|ALL}
-
<bool>
Bool
{0|1|ON|OFF}
0
Remarks
•
Before enabling the channel output, please make sure that the current setting
will not affect the devices connected to the power supply.
•
If <source> is omitted, the command executes the corresponding operation on
the current channel.
Return Format
The query returns 1 or 0.
Examples
:OUTP CH1,ON /*Enables the output for CH1.*/
:OUTP? CH1 /*Queries whether the CH1 output is enabled. The query
returns 1*/
4.8.16
:OUTPut:TRACk[:STATe]
Syntax
:OUTPut:TRACk[:STATe]
<
bool>
:OUTPut:TRACk[:STATe]?
Description
Sets or queries the on/off state of tracking function.
Command System
56
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.