Remarks
•
When the instrument operates in remote lock mode, all the keys on the front
panel cannot be used except for the output on/off key of each channel ,
,
and the power switch key
. At this point, you can press the output on/off
key only to disable channel output.
•
If <bool> is omitted, the command enables the remote lock.
•
to enable or disable remote lock.
Return Format
The query returns 1 or 0.
Examples
:SYST:RWL ON /*Enables the remote lock.*/
4.11.16 :SYSTem:SAMPling
Syntax
:SYSTem:SAMPling
<
mode>
:SYSTem:SAMPling?
Description
Sets or queries the sampling mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{AUTO|HIGH|LOW}
AUTO
Remarks
•
DP2000 series power supply provides three current sampling modes for CH1 and
CH2. You can select “High Curr”, “Low Curr”, or “AUTO”, and the default is
“AUTO”.
-
HIGH: The power supply uses the high range current mode.
-
LOW: The maximum current that can be programmed is 0.5 A and the
maximum current that can be read is 11 mA. When 11 mA is reached, the
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
95
Summary of Contents for DP2000 Series
Page 1: ......
Page 137: ...d Exit Programming Examples Copyright RIGOL TECHNOLOGIES CO LTD All rights reserved 129...
Page 145: ......