RIGOL
DG1000Z Performance Verification Guide
16
Impedance (typical)
50Ω, AC coupling
Sync Output
Level
TTL-compatible
Impedance
50Ω, nominal value
Overvoltage Protection
Overvoltage protection will take effect once any of the following two conditions is
met:
The amplitude setting in the generator is greater than 2Vpp or the output
offset is greater than |2V
DC
|, the input voltage is greater than
±11.5×(1±5%)V (<10kHz).
The amplitude setting in the generator is lower than or equal to 2Vpp or the
output offset is lower than or equal to |2V
DC
|, the input voltage is greater than
±3.5×(1±5%)V (<10kHz).
General Specifications
Power
Power Voltage
100V to 240V (45Hz to 440Hz)
Power Consumption
Less than 40W
Fuse
250V, T3.15A
Display
Type
3-inch TFT LCD
Resolution
320 Horizontal×RGB×240 Vertical Resolution
Color
16M color
Environment
Temperature Range
Operating: 0
℃
to 50
℃
Non-Operating: -40
℃
to 70
℃
Cooling Method
Cooling by fans compulsively
Humidity Range
Less than 30
℃
: ≤95% Relative Humidity (RH)
30
℃
to 40
℃
: ≤75% Relative Humidity (RH)
40
℃
to 50
℃
: ≤45% Relative Humidity (RH)
Altitude
Operating: Less than 3000 meters
Non-Operating: Less than 15,000 meters