Chapter 2 Performance Verification Test
RIGOL
DG1000Z Performance Verification Guide
2-3
5.
Record the reading of the frequency counter and judge whether the reading is
between 0.999 999MHz and 1.000 001MHz.
6.
Set CH1 of DG1000Z to output square, ramp and pulse waveforms (the
frequencies are 1MHz and the amplitudes are 1Vpp) respectively. Record the
readings of the frequency counter respectively and judge whether the readings
are between 0.999 999MHz and 1.000 001MHz.
7.
Repeat steps 1 to 6 to test the frequency accuracy of CH2 and record the test
results.
Test Record Form:
Waveform
Setting
Value
Measurement
Value
Specification
Pass/Fail
Sine
Frequency:
1MHz
Amplitude:
1Vpp
0.999 999MHz to
1.000 001MHz
Square
Ramp
Pulse