DG1000Z Performance Verification Guide
14
the carrier waveform
Sweep Time
1ms to 500s
Hold/Return Time
0ms to 500s
Trigger Source
Internal, External or Manual
Mark
Falling edge of the Sync signal (programmable)
Counter
Function
Frequency, Period, Positive/Negative Pulse Width,
Duty Cycle
Frequency Resolution 7 digits/second (Gate Time =1s)
Frequency Range
1μHz to 200MHz
Period Measurement
Measurement Range 5ns to 16 days
Voltage Range and Sensitivity (Not modulation signal)
DC Coupling
DC Offset Range
±1.5Vdc
1μHz to 100MHz
50mVRMS to ±2.5Vac+dc
100MHz to 200MHz 100mVRMS to ±2.5Vac+dc
AC Coupling
1μHz to 100MHz
50mVRMS to ±2.5Vpp
100MHz to 200MHz 100mVRMS to ±2.5Vpp
Pulse Width and Duty Cycle Measurement
Frequency/Amplitude
Range
1μHz to 25MHz
50mVRMS to
±2.5Vac+dc
DC Coupling
Pulse Width
Minimum
≥20ns
Resolution
5ns
Duty Cycle
Range (Display)
0% to 100%
Input Characteristics
Input Signal Range
Breakdown Voltage
±7Vac+dc
Impedance=
1MΩ
Input Adjustment
Coupling
AC
DC
HF Suppression
ON: input bandwidth=250kHz;
OFF: input bandwidth=200MHz
Input Trigger
Trigger Level Range -2.5V to +2.5V
Trigger Sensitivity
Range
0% (about 140mV hysteresis
voltage) to 100% (about 2mV
hysteresis voltage)