RIGOL
DG1000Z Performance Verification Guide
1
Appendix
Appendix A: Test Result Record Form
RIGOL
DG1000Z Series Function/Arbitrary Waveform Generator
Performance Verification Test Record Form
Model:
Tested by:
Test Date:
Channel: CH1
Frequency Accuracy Test
Waveform
Setting
Value
Measurement
Value
Specification
Pass/Fail
Sine
Frequency:
1MHz
Amplitude:
1Vpp
0.999 999MHz to
1.000 001MHz
Square
Ramp
Pulse
AC Amplitude Accuracy Test
Amplitude
Setting
Value
Setting
Measurement
Value
Specification
Pass/Fail
20mVpp
Frequency:
1kHz
Offset: 0V
DC
Impedance:
50Ω
6.6mVrms to 7.5mVrms
100mVpp
34.7mVrms to 36.1mVrms
500mVpp
174.7mVrms to 178.9mVrms
1Vpp
349.7mVrms to 357.5mVrms
5Vpp
1.75Vrms to 1.7861Vrms
10Vpp
[1]
3.5Vrms to 3.5717Vrms
Note
[1]
: Only applicable to DG1062Z.