4.4
CABLE POWER SUPPLY / PRINTER REF
5010-077
RIELE BERLIN
Photometer 5010 V5+
4.4.1
29.08.2007
Function Description
On devices previous to SN 8002, after a certain time
of operation, may occur error 52 (see chapter 7) due
to the printer controller power supply plug connection
(P23).
To prevent this, it is possible to replace P23 with the
new power supply cable described in this chapter.
The printer controller power supply CN2 will be
connected then directly to the AC / DC power supply,
eliminating the connection through the motherboard.
Beginning with SN 8002 all devices are equipped
with this cable at factory, thus no replacement is
required.
Location
The printer controller cable supply should be placed
as shown here.