3.3 SERVICE
TOOLS
S15
ADC Counts (Dark Adjustment)
The ADC values of the dark adjustment can be read.
The display shows the following information:
•
a D in case of a preselected flow through cuvette
adaptor
•
an E in case of a preselected standard cuvette
adaptor
•
the ADC values assigned to the amplification levels
V0 up to V7
With [E/D] the display shows alternately the values in
combination with the different adaptors E or D.
A D C C O U N T S ( D A R K A D J . )
E
0 : 1 8 4 2 5 1 : 1 8 4 2 8
2 : 1 8 4 3 3 3 : 1 8 4 4 2
4 : 1 8 4 5 7 5 : 1 8 4 8 3
6 : 1 8 5 3 4 7 : 1 8 6 3 0
E X I T E / D
S16 Boost
Wavelength
To improve the low light gain at small wavelengths, the
measurement time is extended when operating below a
definite wavelength. This definite wavelength will be
displayed on the right upper corner of the window.
The default value is 390 nm.
The input has to be confirmed with [E].
B O O S T W A V E L E N G 3 9 0 n m
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN
Photometer 5010 V5+
3.3.11
11.03.2009