4.11 FILTER
WHEEL
RIELE BERLIN
Photometer 5010 V5+
4.11.5
09.04.2009
1.
Run UTILITIES – SERVICE TOOLS - SET WAVELENGTH
(see chapter 3.3).
The assignment of wavelength and filter
position can be done.
The current filter position is blinking.
With [SELECT] the filter position can be set.
With [NEW] the wavelength can be edited.
In the first line the current filter position is
shown.
The new value of the wavelength has to be
entered.
The input has to be confirmed with [E].
2.
Run UTILITIES – SERVICE TOOLS - ADC COUNTS (OPTIC).
Indicated is the current value of the optical
analogue-digital converter. The value is
proportional to the light-current depending on
the selected amplification level and the boost
setting.
To a key actuation the system reacts
possibly only after three seconds.
The functions V0 to V7 set the amplification
level. The function BO increases (boost ON)
or reduces (boost OFF) the time slot. The
functions F0 to F9 place the filter wheel into
the positions 0 to 9. The position 0 corre-
sponds to the filter wheel position at the dark
level adjustment.
Select amplification level V0 and boost OFF.
Stop function by [ESC].
3.
The ADC counts of the new wavelength should be within a range of 15.000 and 600.000.
4.
Finish the operation with [ESC].
S E T W A V E L E N G T H
1 : 3 4 0 6 : 6 2 3
2 : 4 0 5 7 : 9 9 9
3 : 4 9 2 8 : 9 9 9
4 : 5 4 6 9 : 9 9 9
5 : 5 7 8
E X I T N E W S E L E C T
N E W W A V E L E N G T H 1
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S E R V I C E T O O L S P A G E 4 / 5
A D C C O U N T S ( O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 2 1 3 3 6
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B O .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C