4.12 HALOGEN LAMP REF
0554871001
RIELE BERLIN
Photometer 5010 V5+
4.12.4
12.05.2007
Installation
1.
For installation proceed in reverse order.
2.
In case of working with the flow-through cuvette adaptor the cell must be filled with distilled water. In
case of working with the standard cuvette adaptor the cuvette compartment must be empty.
3.
Switch on the Photometer
5010 V5+
.
4.
Check the voltage 12 V
DC
of halogen lamp (see chapter 4.40.4).
5.
Run UTILITIES – ADC COUNTS (OPTIC).
Indicated is the current value of the optical
analogue-digital converter. The value is
proportional to the light-current depending on
the selected amplification level and the boost
setting.
To a key actuation the system reacts
possibly only after three seconds.
The functions V0 to V7 set the amplification
level. The function BO increases (boost ON)
or reduces (boost OFF) the time slot. The
functions F0 to F9 place the filter wheel into
the positions 0 to 9. The position 0 corre-
sponds to the filter wheel position at the dark
level adjustment.
Select amplification level V0 and boost OFF.
Stop function by [ESC].
6.
Check one after another the used positions of the filter wheel.
7.
The ADC counts should be within a range of 20.000 and 600.000
(see chapter 4.40.6 – Adjustment of amplifier stage).
8.
Finish the operation with [ESC].