![Rieber navioven Operating Instructions Manual Download Page 47](http://html.mh-extra.com/html/rieber/navioven/navioven_operating-instructions-manual_1459151047.webp)
Fault - What now?
Rieber GmbH & Co. KG
47
9
Fault - What now?
DANGER
Danger of electric shock. Dangerous electrical voltage.
►
Disconnect appliance from mains before any repair work.
►
Have all electrical repairs carried out by an authorised electrician only.
Repairs during the warranty period may only be carried out by the manufacturer's service department.
If there is a fault, F1 is shown in the display.
Fault
Possible cause
Possible remedy
F1 in the display
Malfunction
Call Rieber Service or
contractual partner
#
User
Display is blank
Electrical connector / electrical
cable defective
Disconnect connecting
plug and lock appliance.
#
Qualified electrician
Appliance does not heat up/
Temperature too high
Heating defective
Machine parameter P entered
wrongly
Controller defective
Replace
#
User
Repair
#
Rieber Service or
authorized partner
Control unit
Not functioning
Control unit defective
Repair
#
Rieber Service or
authorized partner
Humidification does not
work
No water in the water
container
Scale or dirty
Pour drinking water into
the water container
#
User
Descale, clean
#
User
Residual water is not
pumped out
Water pump defective
Repair
#
Rieber Service or
authorized partner