TA Plus Balance
8.2 Error Messages
Error Message/
Error Code
Cause
Solution
OVER ERROR
Weight exceeds maximum capacity
Split load into several smaller loads and weigh them
Replace the tare with a lighter one
Result exceeds the maximum display digit
Clear the calculation result, if the calculation results exceed the
maximum display digit it will continue to do so
UNDER ERROR The load is below the lower limit
Check the position of the weigh pan and reset if necessary
Check for contact with other objects; use the included pan base
only
DISPLAY ERROR The result exceeds the maximum display digit
If calculation results exceed the maximum display digit, it will
continue to do so unless something is changed
LOWER ERROR The specimen weight/reference weight in
Counting/Percentage Mode is below the lower limit
Ensure the specimen weight/reference weight is higher than the
lower limit
ERR001
ERR099
System error
Record the error code and notify the dealer or Rice Lake Weigh
-
ing Systems
ERR703
Operation key pushed when unit is in standby
Hardware issue
Ensure operation key is not pushed when unit is in standby or
starting up from standby
Record the error code and notify the dealer or Rice Lake Weigh
-
ing Systems
ERR705
Initial zero adjustment error
Initial zero adjustment was not completed during startup due to the
unstable load
Ensure weigh pan is properly placed
Check for contact with other objects
Check for wind or vibration
ERR706
Load is out of initial zero adjustment range
Ensure load pan is empty when balance is powered on
ERR709
ERR710
ERR711
Load is unstable at zero adjustment/tare subtraction
Span adjustment time-out error
Improper setting of the weighing pan or pan base is suspected
Check for contact with other objects
Check for wind or vibration
ERR717
Mass of calibration weight differs from designated mass by 1% or
more in external span adjustment
Check the calibration value of the weight and use the proper
calibration weight
ERR718
The mass of the calibration weight is under 50% of the maximum
capacity at span adjustment or internal span adjustment weight
adjustment by external calibration weight
Use a calibration weight which is equal to the maximum capac
-
ity
ERR719
Adjust value by external span adjustment or internal span adjustment
is over 1% of maximum capacity
Execute a
637 REF CAL RESTORE
, then execute an internal
span
adjustment
Check the mass of the weight used for the external span adjust
-
ment
Perform a
636 REF CAL
ERR722
Tare key is pushed during the preset tare operation
Ensure Tare key is not pushed during preset tare operation
ERR723
Out of Zero adjustment range (1.5% of maximum capacity)
Ensure weigh pan is empty while performing a zero adjustment
ERR724
Out of Tare subtraction range (0 g to maximum capacity)
Ensure tare weight is within the tare subtraction range
ERR734
Weight of the sample is out of the importing range at actual value
setting method at Percentage Mode (lower limit to maximum capac
-
ity)
Load the sample of which weight is within the importing range
ERR735
Time-out error of importing the sample weight in the actual value set
-
ting method at Percentage Mode
Improper setting of the weighing pan or pan base is suspected
Check for contact with other object
Check for any wind or vibration
ERR736
The setting value is out of the setting range at numeric value setting
method at Percentage Mode (lower limit to maximum capacity)
Set the value within the range
Table 8-1. Error Codes