![Rice Lake TA Plus 124 Operation Manual Download Page 14](http://html1.mh-extra.com/html/rice-lake/ta-plus-124/ta-plus-124_operation-manual_1454254014.webp)
8
Visit our website
TA Plus Balance
3.3 Precautions Related to a Specimen
Static Electricity
• Synthetic resin- and glass-made specimens are easily charged electrically
• Weighing electrically charged specimens makes the displayed value unstable, reducing the reproducibility of a test result
• Neutralize electrically charged specimens before measurement
Magnetism
• Specimens affected by magnetism weigh differently depending on the position on the pan, reducing reproducibility
• Eliminate the magnetism or place the specimens on a container/plate to protect the weighing mechanism
Moisture Absorption/Evaporation
• Wet or volatile specimens can cause the displayed value to continuously fluctuate due to evaporation
• Put these specimens in a container with a tight fitting lid to prevent evaporation during the weighing process
Specimen Temperature
• Difference in temperature between specimens and the windshield interior generates convection flow, causing errors
• If specimen temperature is excessively high or low, allow its temperature to stabilize at room temperature prior to measuring
• Ensure the windshield interior temperature is equal to room temperature prior to measuring
• Body temperature of operator affects measurement results; handle specimens with tweezers and refrain from putting
hands in the windshield during operation
3.4 Display Navigation
Use the F-keys to navigate through the menus and enter numerical values.
•
= F1: Increments numeric values. Scrolls through menus/selections
•
= F2: Decrements numeric value. Scrolls through menus/selections
•
= F3: Select digits and returns to previous level
•
= F4: Enters the value or a level
3.5 Power On/Off
When balance is plugged in,
displays indicating it is in standby mode.
Press
. A self-check runs, then
000000
displays
.
• Do not press any buttons during the self-check
• The balance starts in the last operation mode set
Press and hold
to put the balance into standby mode (approximately 2-3 seconds).
displays.
3.5.1 Balance Operation Check
Press on the weighing pan lightly to see if the weight display changes.