TA Plus Balance
4.2.3 Calibrating the Internal Weight
This function is used to calibrate the internal weight using an external weight.
To calibrate more accurately use a weight that is equivalent to the weighing capacity.
1. Press
to change the mode.
2. Press
to scroll to
Maintenance
.
3. Press
636 REF CAL
displays.
4. Press
to start the span test.
on 0
and
processing
displays. On completion of the zero point, the display
changes to
On FS.
5. Place a weight in the center of the weighing pan and
on FS
displays.
6.
on O
displays upon completion of the span adjustment.
7. Remove weight from the balance and a blinking
on 0
displays followed by
Ch 0
and
Ch FS
.
8. Upon completion of the calibrating of the internal weight,
636 REF CAL
displays.
9. Press
to exit the setting menu.
4.2.4 Restore the Internal Weight Calibration Value to Default
1. Press
to change the mode.
2. Press
to scroll to
Maintenance
.
3. Press
637 REF CAL RESTORE
displays. The internal weight calibration value is restored to the default
factory settings.
4. Press
and
637 REF RESTORE
displays.
5. Press
to exit the setting menu.
4.2.5 Advice CAL and Automatic Span Adjustment
The TA Plus balance generates an alarm when calibration is needed.
1. Press
to change the mode.
2. Press
to scroll to
Maintenance
.
3. Press
638 ADVICE CAL
displays.
Note