Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
servicing or cleaning.
Figure 2-3. Load Cell Protection
8. When the interface cable is protected and in its final position, complete connections to the indicator. If
necessary, trim corners as described inSection 3.2 on page 5.
9. Check all strain relief fittings for tightness.
10. Put the cover back onto the junction box assembly and slide the junction box back into the floor scale
cutout.
11. Secure the Summit SB-1000 side plate.
2.6
Access Ramps
Access ramps for Summit SB-1000 Bariatric Wheelchair Scales are designed to bolt to the floor, with built-in
mounting plates that attach to the scale feet. When used with access ramps, side movement of the scale is
automatically eliminated, and no other mounting plates are necessary.
The access ramp can only be attached to the scale on one side of the scale which is opposite the junction box. For
example, the scale shown in Figure 2-4 could have an access ramp on the left side as shown which sits opposite
the junction box. The top and bottom sides, however, will not accept the ramp mounting plates.
Figure 2-4. Access Ramp
SCALE FLOOR
LOAD CELL CABLE
Longitudinal axis
of load cells
Junction box
location