Introduction
1
1.0
Introduction
This manual is intended for use by service technicians responsible for installing and servicing the
IQ plus
2100
digital bench scale.
Configuration and calibration of the indicator can be accomplished using the indicator front panel keys, the EDP
command set, or Version 2.3 or later of the Revolution
™
configuration utility. See Section 3.1 on page 17 for
information about configuration methods.
Some procedures described in this manual require work inside the indicator enclosure. These procedures
are to be performed by qualified service personnel only.
This manual can be viewed or downloaded from the Rice Lake Weighing Systemswebsite at
www.ricelake.com/manuals
Warranty information can be found on the website at
1.1
Overview
The
IQ plus 2100
digital bench scale consists of an
IQ plus 210
digital weight indicator and an
RL2100
bench
scale. The indicator can be mounted to the bench scale using the 12
”
column or the attachment bracket, or can be
separately mounted on the tilt stand.
The
IQ plus 210
is a single-channel digital weight indicator housed in a NEMA 4X/IP66-rated stainless steel
enclosure. The indicator front panel consists of a large (.8 in, 20 mm), six-digit, seven-segment LED display, seven
LED annunciators, and two piezo switches used to zero the scale and select the displayed units. Features include:
• Drives up to four 350
load cells
• Supports 4- and 6-wire load cell connections
• Two configurable digital inputs
• Single serial port supports full duplex, RS-232 communications at up to 9600 bps
• Available in 115 VAC and 230 VAC versions
The RL2100 bench scale is available in the following configurations:
• 5, 10, and 25 lb (2.5, 5, and 10 kg) capacities with a 10
”
x 10
”
(25 cm x 25 cm) platform
• 50 and 100 lb (25 and 50 kg) capacities with a 12
”
x 12
”
(30 cm x 30 cm) platform
A single point anodized aluminum load cell is standard in the
RL2100
bench scale. A stainless steel clamshell
enclosure is also available for load cell protection.
1.2
Operating Modes
The
IQ plus 210
has two modes of operation:
Weigh mode
Weigh mode is the production mode of the indicator. The indicator displays the gross weight, and the LED
annunciators described on page 2 indicate scale status and the type of weight value displayed. Once
configuration is complete and a legal seal is affixed to the back of the indicator, this is the only mode in which
the
IQ plus 210
can operate.
Setup mode
Most of the procedures described in this manual require the indicator to be in setup mode, including
configuration and calibration.
To enter setup mode, remove the large fillister head screw from the enclosure backplate. Insert a screwdriver or
a similar tool into the access hole and press the setup switch once. The indicator display changes to show the
word
CONFIG
.
WARNING