![Rice Lake 650-10-1 Technical Manual Download Page 8](http://html1.mh-extra.com/html/rice-lake/650-10-1/650-10-1_technical-manual_1454107008.webp)
4
RL-650 Series Neonatal Scale
1.5
Operation
1.5.1
Before Using the Scale
Use the following steps for initial setup.
1. Press
to turn on scale. After self test,
0
displays.
2. Place weight (maximum 33 lb/15 kg) on the scale.
flashes on the displays until weight calculation is
complete.
The weight calculation is complete once the motion indicator turns on.
3. Press
to select the weight mode, lb or kg.
4. Press
. Remove the weight, the weight continues to display.
5. Press
to remove hold.
0
displays.
If the setup procedure failed, see
Section 6.1 on page 15
. If the problem is not resolved, contact a qualified
service provider.
1.5.2
Weighing
Use the following steps to weigh on the scale.
1. Press
to turn the scale on. After self test,
0
displays.
2. Place a pad or other accessory to be used on the weighing tray
.
3. Press
to remove the weight of a pad or other accessory,
0
displays.
4. Place the infant on the scale. Infants weight displays.
5. Press
to weigh the infant again, for more accurate results.
6. Remove the infant from the scale.
0
displays.
If the scale is not used for two minutes, it turns off.
1.5.3
Hold and Release Function
Use the following steps to use the
HOLD/RELEASE
function.
1. Press
to turn the scale on. After self test,
0
displays.
2. Place the infant on the scale. Infants weight displays.
3. Press .
4. Remove infant from the scale, the weight remains on the scale and
HOLD
displays.
5. Press
to return to zero.
can be pressed with infant still on the scale. Once infant is removed,
0
displays.
-----
-----
UNITS
CLEAR
HOLD
RELEASE
HOLD
RELEASE
Note
TARE
REWEIGH
Note
HOLD
RELEASE
HOLD
RELEASE
Note
HOLD
RELEASE
Summary of Contents for 650-10-1
Page 1: ...RL 650 Neonatal Scale 650 10 1 Technical Manual PN 172982 Rev C...
Page 2: ......
Page 22: ...18 RL 650 Series Neonatal Scale...
Page 23: ......