8
RL-650 Series Neonatal Scale
3.0
Scale Configuration
Use the scale configuration to set up the scales parameters and options, that are essential for the functioning of the scale.
Access to the setup switch is located on the underside of the unit. Use a non-conductive object and gently press the
internal
setup switch.
Figure 3-1. Setup Switch Location
3.1
Maintenance Menu Parameters
The measuring unit of the menu will be according to the measuring unit of the last calibration.
3.1.1
User Menu
1. Press
to turn the scale on, when
START
displays press and hold
until
ID
and firmware ID number
flashes on the display.
2. Press .
TEST
displays.
3. Press
to enter menu or
to scroll to next parameter.
Parameter
Settings
Description
TEST
Test menu
RID xxxxx
Internal ID number
VER xx.xx
Development internal sub version
A2D
Raw weight data
VALUE
Processed weight in live display
BAT
Raw bat value[BAT/16]
USER
Access some of the parameters from the PROG menu.
BAUD
(96) -
UART baud rate.
Values: 48, 96, 144, 192, 288, 384, 576, 1152
TOFF
(5) -
Auto off timer in minutes (0 = Always on) - maximum 9
When using external power supply this parameter is irrelevant
LIVE
(OFF)
OP0 – Live weighing (0=Disable, 1=Enable, Default is Disable).
MAIN
Communication protocol (0=ESC, 1=Maintenance, Default is ESC). Default – OFF
OP1
SPACE (7)
Number of new lines after print (hex: 0x0d 0x0a)
SAVE
Programming done, parameters save to EEPROM.
Table 3-1. User Menu Parameters
ENTER
ENTER
ENTER