Rice Lake 650-10-1 Technical Manual Download Page 14

10

RL-650 Series Neonatal Scale

* The lb menu values will be multiplied by 2. (example: full 250.0 kg = 500.0 lb)

* The decimal point location is fixed in this program to 0.000

ARW 

0.050 kg\0.100 lb

Auto-Reweigh – restarts weigh algorithm if weight changes by more than set value; 
max full capacity* 

SAL 

0.005 Sec

Semi-Auto-Live – interval between weight displays during algorithm process; max 0.9 sec

ROUND 

0.005 kg\0.010 lb

Scale resolution – Displays the high round if in dual range

• values in kg: 1, 1, 2, 5,   10, 20, 50,   100
• values in lb:  1, 2, 5, 10, 20, 50, 100, 200

The decimal point location is set according to DISP parameter

BAUD 

96

UART baud rate – settings: 48, 96, 144, 192, 288, 384, 576, 1152
Value entered is multiplied by 100  

example: 96 x 100 = 9600

ATOL 

2

0-255

Algorithm Initial Tolerance – if value is above 255, doesn’t proceed and returns to previous 
value

ALEN 8

0-10

Algorithm Initial Exponent – if value is above 10, it returns to previous value

ATOUT 11

0-15

Algorithm Maximal Exponent – if value is above 15, it returns to previous value.

TOFF 

5

Auto Off Timer 0-9 minutes (0 = always on) only applies when battery powered

TLOC

3

Auto Release Lock Timer – in seconds (0 = NO RELEASE, can have values from 0 to 9)
Every x seconds the lock is released and the weighing algorithm restarts.

LANG

1

Print Language – 0 = English, 1 = French

SPACE

7

Number of new lines after print (hex: 0x0d 0x0a)

UNITS 

kg\lb

Units of Measure displayed (kg, lb or kg\lb)

OP

--

Binary options:

• OP0 – Live weighing (0=Disable, 1=Enable, Default is Disable).
• OP1 – Communication protocol (0=ESC, 1=Maintenance, Default is ESC).
• OP2  –  Reserved
• OP3 – RTC power (0=Disable, 1= Enable, Default is Enable).
• OP4 – SEMI-AUTO-LIVE (0=Disable, 1= Enable, Default is Enable).
• OP5 –full calculation (0= Spatial, 1=multiply by 2, Default is multiply by 2).
• OP6 – TARE (0=Disable, 1=Enable, Default is Enable).
• OP7 – BAT type (0=Dry batteries, 1= rechargeable batteries, Default is rechargeable).
• OP8 – OIML Mode (0=Disable, 1=Enable, Default is Enable).
• OP9 – USB boost (0= Disable, 1=  Enable, Enable).
• OP10 – OZ display (0= Disable, 1= Enable, Disable).
• OP11 – Gravity Update - GUPDAT (0= Disable, 1= Enable, Disable).
• OP12 – Enable Print function - (0= Disable, 1= Enable, Disable).

SAVE

--

Press to save parameters to EEPROM (only displays if changes have been made)

DONE

--

Scale reboots

Parameter

Default

Description and Settings

Table 3-4. Programming Menu Parameters (Continued)

Summary of Contents for 650-10-1

Page 1: ...RL 650 Neonatal Scale 650 10 1 Technical Manual PN 172982 Rev C...

Page 2: ......

Page 3: ...Parameters 8 3 1 1 User Menu 8 5 0 RS 232 Communication 12 5 1 Pushbutton Keypad Print 12 5 2 USB Connection 12 6 0 Troubleshooting and Maintenance 15 6 1 Troubleshooting Testing 15 6 2 1 Basic Maint...

Page 4: ...ii RL 650 10 1 Technical Manual Contents Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Page 5: ...baby and damage to the scale See Section 1 2 on page 2 Follow all instructions for installation and usage included in this manual The manufacturer is not responsible for any damage or injury from inc...

Page 6: ...to determine weight of babies neonates It is not intended to diagnose prevent and monitor diseases To avoid cross contamination the scale plate must be cleaned after each use Avoid direct skin contac...

Page 7: ...s on display after removing infant from scale 10 On Off Turns scale on off 11 Reweigh Allows repeated reweighing of an infant a long key press activates the print function 12 Hold Release First press...

Page 8: ...scale on After self test 0 displays 2 Place a pad or other accessory to be used on the weighing tray 3 Press to remove the weight of a pad or other accessory 0 displays 4 Place the infant on the scal...

Page 9: ...weight and NET displays 5 Remove the infant and item tared The tared weight is stored in memory 6 To cancel the tare weight press until display returns to 0 and GROSS displays The tare weight cancels...

Page 10: ...tery operation when an AC power source is not available 2 2 1 Install Batteries 1 Set the scale on a sturdy flat surface 2 Tip the unit on its side to access the battery chamber Figure 2 1 Battery Cha...

Page 11: ...cale Setup 1 Ensure the scale is placed on a sturdy flat surface 2 Remove the transport support tabs from each corner of the scale base Figure 2 3 Support Tab Corners 3 The scale is equipped with four...

Page 12: ...lay 2 Press TEST displays 3 Press to enter menu or to scroll to next parameter Parameter Settings Description TEST Test menu RID xxxxx Internal ID number VER xx xx Development internal sub version A2D...

Page 13: ...ngs MONT Set the next verification date in month Default 11 If set to zero no verification message CNTL will be displayed On entering this menu will always show 11 month The CNTL message is displayed...

Page 14: ...tes 0 always on only applies when battery powered TLOC 3 Auto Release Lock Timer in seconds 0 NO RELEASE can have values from 0 to 9 Every x seconds the lock is released and the weighing algorithm res...

Page 15: ...an those listed in Step 5 press again and the far right digit flashes Press the up or down arrows to change the weight value The left arrow moves to the next digit on the left allowing changes to the...

Page 16: ...eight was either in motion or out of range The print will show on the display when it is transmitting data If displaying weight the scale will send out the following 21 character string xxxxxxxxx SP u...

Page 17: ...Driver Click on Download to open and download the driver to a local computer 4 When the USB cable is connected to the indicator and the scale is turned on a display prompts to navigate through the so...

Page 18: ...terminal emulation program such as Hyperterminal via the USB driver Select the port assigned to the software driver Parpar to establish a port The terminal emulation program is necessary to view info...

Page 19: ...d Dry with a clean soft cloth Do not immerse the scale into cleaning or other liquid solutions Do not use Isopropyl alcohol or other solutions to clean the display surface Symptom Possible Cause Corre...

Page 20: ...er is UL and C UL approved Battery Type AA Lithium Ion disposable Nominal Voltage 1 5 V Rated capacity 2900 mAh Working voltage 1 30V 200 mA discharge Environmental Operating temperature 50 to 95 F 10...

Page 21: ...caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modifie...

Page 22: ...18 RL 650 Series Neonatal Scale...

Page 23: ......

Page 24: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Speci...

Reviews: