10
RL-650 Series Neonatal Scale
* The lb menu values will be multiplied by 2. (example: full 250.0 kg = 500.0 lb)
* The decimal point location is fixed in this program to 0.000
ARW
0.050 kg\0.100 lb
Auto-Reweigh – restarts weigh algorithm if weight changes by more than set value;
max full capacity*
SAL
0.005 Sec
Semi-Auto-Live – interval between weight displays during algorithm process; max 0.9 sec
ROUND
0.005 kg\0.010 lb
Scale resolution – Displays the high round if in dual range
• values in kg: 1, 1, 2, 5, 10, 20, 50, 100
• values in lb: 1, 2, 5, 10, 20, 50, 100, 200
The decimal point location is set according to DISP parameter
BAUD
96
UART baud rate – settings: 48, 96, 144, 192, 288, 384, 576, 1152
Value entered is multiplied by 100
example: 96 x 100 = 9600
ATOL
2
0-255
Algorithm Initial Tolerance – if value is above 255, doesn’t proceed and returns to previous
value
ALEN 8
0-10
Algorithm Initial Exponent – if value is above 10, it returns to previous value
ATOUT 11
0-15
Algorithm Maximal Exponent – if value is above 15, it returns to previous value.
TOFF
5
Auto Off Timer 0-9 minutes (0 = always on) only applies when battery powered
TLOC
3
Auto Release Lock Timer – in seconds (0 = NO RELEASE, can have values from 0 to 9)
Every x seconds the lock is released and the weighing algorithm restarts.
LANG
1
Print Language – 0 = English, 1 = French
SPACE
7
Number of new lines after print (hex: 0x0d 0x0a)
UNITS
kg\lb
Units of Measure displayed (kg, lb or kg\lb)
OP
--
Binary options:
• OP0 – Live weighing (0=Disable, 1=Enable, Default is Disable).
• OP1 – Communication protocol (0=ESC, 1=Maintenance, Default is ESC).
• OP2 – Reserved
• OP3 – RTC power (0=Disable, 1= Enable, Default is Enable).
• OP4 – SEMI-AUTO-LIVE (0=Disable, 1= Enable, Default is Enable).
• OP5 –full calculation (0= Spatial, 1=multiply by 2, Default is multiply by 2).
• OP6 – TARE (0=Disable, 1=Enable, Default is Enable).
• OP7 – BAT type (0=Dry batteries, 1= rechargeable batteries, Default is rechargeable).
• OP8 – OIML Mode (0=Disable, 1=Enable, Default is Enable).
• OP9 – USB boost (0= Disable, 1= Enable, Enable).
• OP10 – OZ display (0= Disable, 1= Enable, Disable).
• OP11 – Gravity Update - GUPDAT (0= Disable, 1= Enable, Disable).
• OP12 – Enable Print function - (0= Disable, 1= Enable, Disable).
SAVE
--
Press to save parameters to EEPROM (only displays if changes have been made)
DONE
--
Scale reboots
Parameter
Default
Description and Settings