Rice Lake 482-AG Technical Manual Download Page 22

482-AG Livestock Digital Weight Indicator

18

Visit our website 

www.RiceLake.com

3.1 Configuration Menu 

Figure 3-1. Configuration Menu

Parameter

Settings

Description

GRADS

4000

1 – 100000

Graduations – specifies the number of full scale graduations; the value entered must be in the range of 1–100,000 and 

should be consistent with legal requirements and environmental limits on system resolution

ZTRKBN

0.6 dd

0.0 – 3.0dd

Zero track band – automatically zeros the scale when within the range specified, as long as the input is within the configured 

zero range (ZRANGE parameter); selections are ± display divisions; maximum legal value varies depending on local 

regulations

ZRANGE

2%

1%–100%

Zero range – selects the range within which the scale can be zeroed; the 2% selection is ± 2% around the

calibrated zero point, for a total range of 4%; indicator must be at standstill to zero the scale

INIZR

0%

0–100%

Initial zero range at power up; see Error messages HINOFF and LINOFF in 

Section 8.1 on page 48

; if on power up, the 

weight value is between the ±% range specified of Calibrated Zero, the indicator will zero off that weight; if outside that 

range, the indicator will display HINOFF or LINOFF, depending on whether the initial reading is above or below the range

NOTE: If not set at zero, hopper/scale MUST be empty when powered up.

MOTBAN

1.0d

0.0d–100d

Motion band – sets the level, in display divisions, at which scale motion is detected; if motion is not detected, the standstill 

symbol lights; some operations, including print, tare and zero, require the scale to be at standstill; maximum legal value 

varies depending on local regulations

NOTE: Setting the MOTBAN setting to 0.0 disables motion sensing and the scale will always be

considered stable.

OVRLOA

FS+2%

FS+1D

FS+9D

FS

Overload – determines the point at which the display blanks and an out-of-range error message is displayed; maximum 

legal value varies depending on local regulations

DSPRAT

0.1 SEC

0.25 SEC

0.5 SEC

0.75 SEC

1 SEC

NONE

Display rate – sets the update rate for displayed values; Values are in seconds (SEC)

No display filtering – the display is updated at up to the A/D sample rate

SMPRAT

5HZ

10HZ

20HZ

40HZ

A/D Sample rate – selects measurement rate, in samples per second, of the analog-to-digital converter; lower sample rate 

values provide greater signal noise immunity

DIGFIL

ROLFIL

OFF

ADPFIL

The digital filter setting determines which if any digital filter option is used

• ROLFIL - Rolling Average Filter

• OFF - Disables digital filter

• ADPFIL - Adaptive Filter

ADPFIL

--

Adaptive Filter parameter settings, see 

Section 3.1.1 on page 19

ROLFIL

--

Rolling Average Filter parameter settings, see 

Section 3.1.1 on page 19

DSPMOD

--

Display Mode parameter settings, see 

Section  on page 21

Table 3-2.  Configuration Menu Parameters

CONFIG

FORMAT

CALIBR

PROGRM

PASWRD

TEST

GRADS

ZTRKBN

ZRANGE

MOTBAN

OVRLOA

DSPRAT

SMPRAT

ROLFIL

DIGFIL

ADPFIL

DSPMOD

INIZR

See 

Figure 3-2 on page 19

Summary of Contents for 482-AG

Page 1: ...482 AG Livestock Digital Weight Indicator Version 1 00 Technical Manual PN 167281 Rev B February 14 2020...

Page 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Page 3: ...8 3 Print Ticket 6 1 8 4 Shrink Factor 6 1 8 5 Head Count 7 1 8 6 Load Total 7 2 0 Installation 8 2 1 Unpacking and Assembly 8 2 2 Enclosure Disassembly 8 2 3 Cable Connections 8 2 3 1 Cable Grounding...

Page 4: ...Mode 38 5 2 Live Mode Operation 39 5 3 Hold Mode Operation 39 5 3 1 Hold Settings 39 5 3 2 Weighing Groups in Hold Mode 40 6 0 EDP Commands 41 6 1 The EDP Command Set 41 6 1 1 Key Press Commands 41 6...

Page 5: ...o equipment or corruption to and loss of data General Safety Do not operate or work on this equipment unless this manual has been read and all instructions are understood Failure to follow the instruc...

Page 6: ...ration mode To enter the configuration mode 1 Remove the configuration switch access screw from the back of the enclosure See Figure 2 1 on page 8 2 Insert a non conductive tool into the access hole a...

Page 7: ...b oz tn or t Also used as a scroll left key to navigate to different menus or to toggle to another digit when editing a value Sends an on demand print format out the serial port provided the condition...

Page 8: ...nciator signifies that the indicator is still in the middle of collecting load data The previous group data totals are saved in case of a power down in the middle of collecting load groups 3 Zero Cent...

Page 9: ...ht to view the values for that parameter When the desired value displays press Tare to select the value and move back up one level To edit numerical values use the navigation keys to select the digit...

Page 10: ...ry units The current unit annunciator is displayed Units cannot be changed once a group weighment has been started 1 8 3 Print Ticket The print key sends the group data to the serial port This key wil...

Page 11: ...le is sampled for the set sampling time duration SPTIME When the sample time is complete the average weight of the samples collected is displayed and the current weight units annunciator flashes indic...

Page 12: ...y 4 Lift the backplate away from the enclosure and turn it over to access boards The display cable will still be attached Be careful when lifting and turning over the backplate to prevent the backplat...

Page 13: ...trip the insulation and foil from the cable 1 2 15 mm past the grounding clamp 2 Fold the foil shield back on the cable where the cable passes through the clamp 3 Ensure the silver conductive side of...

Page 14: ...acket Battery and Battery Holder Power Supply CPU and A D Board Battery Option Battery Charger Board 20mA_Out RXD1 TXD1 GND RXD2 TXD2 GND GND Sig SIG SEN SEN EXC EXC Load Cell Connection RS 232 Connec...

Page 15: ...rn off Battery charge time is approximately 16 hours to 100 Figure 2 7 Battery Charging Cycle Example When installing a 482 AG with a battery option it is recommended that the CPU board jumper 4 on th...

Page 16: ...tion bracket if installed will need to be removed first 2 7 Enclosure Reassembly Once cabling is complete position the backplate over the enclosure and reinstall the backplate screws Use the torque pa...

Page 17: ...ctrical contacts while preventing access into Legal for Trade configuration parameters There is an alternative cover over the A D to ensure the A D and load cell connection cannot be changed after the...

Page 18: ...Digital Weight Indicator 14 Visit our website www RiceLake com 2 8 Replacement Parts Figure 2 11 Replacement Parts 4 1 1 2 3 5 7 7 7 8 4 9 19 16 18 17 6 23 25 26 3 20 21 27 15 14 13 22 24 31 30 6 11...

Page 19: ...TCP IP 1 11 131327 Screw 4 40 x 0 188 8 13 170537 Overlay 482_AG membrane 1 14 131343 Enclosure front 1 15 153776 Display board LCD 7 segment 1 16 131316 Power supply 85 265 VAC 1 17 131317 Battery ch...

Page 20: ...7223 15 position screw terminal 1 155230 Ribbon cable CPU to display 1 155231 Ribbon cable CPU to A D 1 155232 Ribbon cable PS to CPU Board 1 155233 Ribbon cable CPU to Option 1 155234 Pwr cord 115 VA...

Page 21: ...switch access screw Menu Required Password Description CONFIG Configuration Config Graduations scale capacity configure zero tracking zero range motion band overload tare function sample rate and dig...

Page 22: ...e level in display divisions at which scale motion is detected if motion is not detected the standstill symbol lights some operations including print tare and zero require the scale to be at standstil...

Page 23: ...the filter from resetting adaptive filtering The threshold will not reset filtering until the specified lock time is met RFSTG1 RFSTG2 RFSTG3 2 4 8 16 32 64 1 Rolling filter stages the number of A D s...

Page 24: ...the previous filtered output value If the difference between the new value and the previous output value is greater than the adaptive filter threshold the adaptive filter is reset to the new weight va...

Page 25: ...G1 3 to eliminate the vibration effects on the scale Leave RFTHRH set to NONE 7 Determine the lowest effect value for the RFSTG1 3 parameters If needed the rolling filter cutout sensitivity RFSENS and...

Page 26: ...y the 10000 lb secondary units value Parameter Choices Description PRIMAR SECNDR DECPNT DSPDIV UNITS Primary Units specifies the capacity decimal position display divisions and units used Secondary Un...

Page 27: ...WVAL Press to display the test weight value The value may be edited as required press to save the edited value WSPAN CAL Press to display ENTER then press to capture the net span value EDIT Press to d...

Page 28: ...he consecutive number start up parameter specifies the initial consecutive number CONSNU value used when the instrument is reset or configured NOTE Consecutive numbering starts with number entered Exa...

Page 29: ...the indicator is awake and will increases battery life The OFF setting stops the indicator from waking up then to wake up the indicator the power key must be pressed The indicator only checks for a de...

Page 30: ...The user password if enabled restricts access to the serial and print formatting menus Once a user password has been entered the existing user password is required before the password can be edited 0...

Page 31: ...ode there is only one number displayed Audit Trail The Audit trail number is incremental for a legally relevant configuration parameter changes and all calibration changes The last date the scale was...

Page 32: ...is set to NONE the indicator operates normally COM 2 will not stream even if the TRIGER is set to a streaming parameter STR1 5 LOCAL When the streaming parameter is set to LOCAL the indicator operate...

Page 33: ...other string is transmitted ECHO ON OFF The echo parameter specifies whether characters received by the port are echoed back to the sending unit PRNMSG OFF ON The print message parameter determines if...

Page 34: ...cter parameter sets the line termination character s for data sent EOLDLY 000 0 255 The end of line delay parameter specifies in 0 1 second intervals the delay between transmitted lines of data NOTE A...

Page 35: ...one of the five fixed formats that follow Rice Lake Stream Data Format Figure 3 10 STR 1 Rice Lake Stream Data Format STX POL wwwwwww UNIT G N S TERM ASCII 02 decimal Polarity space Positive Negative...

Page 36: ...x2 1 1 x5 Even parity bit Requires EDP Port setup Status Word A Status Word C ASCII 20h space Even parity bit Requires EDP Port setup Status Word B 0 Gross Display Mode 1 Net Display Mode 0 Positive...

Page 37: ...unce space none Space Character CR POL wwwwww S SP UNIT SP G N SP SP ETX Space character 2 Space Character Status space valid or invalid m motion o over under range Space Polarity for positive for neg...

Page 38: ...oup Weight format string Default NL2 TD NL Group Number GN NL Head Count HC NL Weight G NL Average Weight LG NL LTFMT Load Total format String Default NL2 TD NL Total Group Nm TGN NL Total Group Weigh...

Page 39: ...ased leakage current from the battery when the indicator is turned off If the battery option is installed and CPU Jumper 4 is disconnected the indicator only turns on and off by pressing the Power key...

Page 40: ...show the test weight value 13 Press to edit value using the procedure shown in Section 1 7 on page 5 to enter the actual value of the test weights to be used for the span calibration Press to store t...

Page 41: ...VAL nnnnnn CR 4 Send the SC WSPAN EDP command to calibrate span The indicator displays CAL while the calibration is in progress 5 To remove an offset value clear all weight from the scale including it...

Page 42: ...will be applied to all of the groups making up the load 2 Load the first group onto scale 3 Enter the head count on the numeric key pad See Section 1 8 5 on page 7 4 Press The group information is se...

Page 43: ...or is reset or the power is cycled 5 3 1 Hold Settings When using the hold mode the following settings should be set prior to weighing see Section on page 21 Set the SPTIME parameter Set the HDELTA pa...

Page 44: ...cleared for example to add additional animals by pressing Then repeat step three and enter the new head count 5 If a reprint of the individual group ticket is required press again to resend the curren...

Page 45: ...mmands and command syntax used for each of these groups 6 1 1 Key Press Commands Key press EDP commands simulate pressing the keys on the front panel of the indicator Most commands can be used in both...

Page 46: ...e value is either a number or a parameter value Use no spaces before or after the equal sign If an incorrect command is typed or the indicator is not in the configuration mode the indicator responds w...

Page 47: ...LB KG OZ TN T G Table 6 4 FORMAT EDP Commands Command Description Values SC WZERO Zero calibration SC WVAL Test weight value value SC WSPAN Span calibration SC REZERO Rezero SC LC CD Read or set dead...

Page 48: ...P port echo ON OFF EDP PRNMSG p Enable print message when a demand print is performed OFF ON EDP STRUR 2 Legal for Trade or industrial stream STRLFT STRIND EDP STREAM Specifies the operating of the in...

Page 49: ...power up nnnnnn DATESET View or set date The response will be in the format and using the separators as defined in configuration to set first query the date then send DATESET in exactly the same form...

Page 50: ...tly displayed on the indicator The 300 character limit of each print format string includes the output field length of the print formatting tokens not the token length For example if the indicator is...

Page 51: ...front panel PFRMAT menu see Figure 3 15 on page 34 for print format menu structure The print format strings may be edited by changing the decimal values of the ASCII characters in the format string So...

Page 52: ...error Weight value is too large to be displayed Gross overload limit Gross value exceeds overload limit check configuration or signal input level overload can be caused by input signal 45 mV or common...

Page 53: ...is run see Table 8 2 for error condition information Example the value yyyyy 50815 represents the decimal equivalent of the binary value 1100 0110 0111 1111 Using the bit assignments in Table 8 2 this...

Page 54: ...25 E 69 45 e 101 65 Ctrl F ACK 06 06 38 26 F 70 46 f 102 66 Ctrl G BEL 07 07 39 27 G 71 47 g 103 67 Ctrl H BS 08 08 40 28 H 72 48 h 104 68 Ctrl I HT 09 09 41 29 I 73 49 i 105 69 Ctrl J LF 10 0A 42 2A...

Page 55: ...EA 139 8B 1 2 171 AB 203 CB 235 EB 140 8C 1 4 172 AC 204 CC 236 EC 141 8D 173 AD 205 CD 237 ED 142 8E 174 AE 206 CE 238 EE 143 8F 175 AF 207 CF 239 EF 144 90 176 B0 208 D0 240 F0 145 91 177 B1 209 D1...

Page 56: ...ceLake com 8 3 Front Panel Display Characters Figure 8 1 provides the seven segment LCD character set used to display alphanumeric characters on the 482 AG front panel Figure 8 1 Display Characters 9...

Page 57: ...Excitation Voltage 5 0 1 VDC Sense Amplifier Differential amplifier with Four and Six wire sensing Analog Signal Input Range Up to 7 mV V Analog Signal Sensitivity 0 1 V graduation minimum 0 5 V grad...

Page 58: ...embossed keys tactile feel Environmental Operating Temperature 10 40 C legal 10 50 C industrial Storage Temperature 25 70 C Humidity 0 95 relative humidity Enclosure Enclosure Dimensions 9 5 x 6 x 2 7...

Page 59: ......

Page 60: ...ada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change with...

Reviews: