482-AG Livestock Digital Weight Indicator
48
Visit our website
8.0
Maintenance and Troubleshooting
8.1 Error Messages
Error message can be viewed on the indicator front panel or from a remote location by using EDP commands.
8.1.1 Displayed Error Messages
The 482-AG provides a number of front panel error messages to assist in problem diagnosis.
lists these messages
and their meanings.
Shorting excitation voltage shuts excitation voltage off. The only way to restore excitation voltage is to cycle power.
Error Messages
Description
Solution
E A/D
A/D physical error
Call Rice Lake Weighing Systems Service at 800-472-6703
EEEROM
EEPROM physical error
EVIREE
Virgin EEPROM
Use TEST menu to perform DEFLT (restore defaults) procedure then recalibrate load cells
EPCKSM
Parameter checksum error
EACKSM
A/D calibration checksum error
A/D converter requires recalibration. Call Rice Lake Weighing Systems Service
EFCKSM
Printer format checksum error
Call Rice Lake Weighing Systems Service at 800-472-6703
ELCKSM
Load cell calibration checksum error Recalibrate load cells
EIDATA
Internal RAM checksum error
Call Rice Lake Weighing Systems Service at 800-472-6703
E REF
A/D reference error
A/D converter requires recalibration; call Rice Lake Weighing Systems Service
ERROR
Internal program error
Check configuration; call Rice Lake Weighing Systems Service if unable to clear error by
cycling power or if error recurs
OVERFL
Overflow error
Weight value is too large to be displayed
========
Gross > overload limit
Gross value exceeds overload limit; check configuration or signal input level;
overload can be caused by input signal > 45 mV or common mode voltage > 950 mV
- - - - - -
Gross < 20d behind zero
Gross value is more than 20 divisions behind zero (OIML)
RNGERR
GRADS > 100,000
WVAL > 100,000
Only displays in configuration mode
EEPERR
EEPROM error
Call Rice Lake Weighing Systems for service at 800-472-6703
HINOFF
High offset
Zero load at startup is more than initial zero range (INIZR) setting of calibration zero–remove
the extra load
LINOFF
Low offset
Zero load at startup is less than initial zero range (INIZR) setting of calibration zero–add the
missing load
NOBATT
No battery
The RTC lost time/date tracking at previous power off state due to low battery or no battery
condition; the printer, accumulator and audit functions will fail to get time and date
NOTE: Refers to internal coin battery only, not the rechargeable battery option.
EPCKSM
Configuration checksum
The checksum value of configuration has changed from that stored in memory; Enter into the
Config menu, then back out; The indicator display Saving; If this does not clear the error,
enter the confiuration menu, navigate to the Test Menu and navigate to Default; Press enter
and the indicator will be reset to default settings
OIMLER
OIML parameter error
Parameter set incorrectly for use in the OIML mode; Check configuration and change as
necessary
Example: Primary units set for lb or oz
EE-ACC
Accumulator error
Error with the accumulator such as attempting to display an accumulated value greater than
six digits
Table 8-1. 482 Error Messages
Note