420 Plus Installation Manual - Configuration
17
3.2.1
Configuration Menu
Figure 3-5. Configuration Menu
CONFIG Menu
Parameter
Choices
Description
Level 2 submenus
GRADS
10000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
To calculate GRADS, use the formula, GRADS =
Capacity
/
Display Divisions
.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBND
0
number
Automatically zeroes the scale when within the range specified, as long as the input is
within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.
For scales using linear calibration, do not set the zero tracking band to
a value greater than that specified for the first linearization point.
ZRANGE
1.900000
number
Selects the range within which the scale can be zeroed. The 1.900000 default value is ±
1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use the default value for legal-for-trade applications.
Table 3-2. Configuration Menu Parameters
XXXXXXX
XXXXXXX
XXXXXXX
XXXXXXX
GRADS
10000
0
number
ZTRKBN
ZRANGE
1.900000
MOTBAN
FS+2%
FS+9D
FS+1D
FS
OVRLOA
1
4
2
8
DIGFL1
32
16
64
number
1
4
2
8
DIGFL2
32
16
64
1
4
2
8
DIGFL3
32
16
64
BOTH
PBTARE
NOT
ARE
KEYED
T AREFN
ALGOUT
DIG IN
SETPNT
PROGRM PFORMT
SERIAL
CALIBR
CONFIG FORMAT
8OUT
32OUT
16OUT
64OUT
DFSENS
DFTHRH
2OUT
128OUT
4OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
15HZ
60HZ
30HZ
7.5HZ
SMPRAT
VERS
number
1
number
Note