Operative Instructions
38
10.5 Alarm table
ALARM
DESCRIPTION
E00
PTC PROBE BREAK
First firmware installation.
E01
PTC PROBE SHORT C
Chamber PT1000 / PT100 probes broken or short-circuited.
E02
NO PTC CONNECTED
No PT1000 / PT100 probes has been enabled (board setting error).
E03
PTC DETECTION
Mix of used probes (PT1000 / PT100).
E04
DETERGENT EMPTY
Detergent liquid tank empty - nozzle probe failure.
E05
SOLVENT EMPTY
Solvent liquid tank empty - nozzle probe failure.
E10
WATER NOT FILLED
Mains tap is closed or the filter is dirty.
E11
LOW W. LEV. OFF
Level probe malfunctioning in the water tank, faulty or disconnected.
E12
DOOR TIMEOUT
Door not open or close (only for automatic door).
E13
DOOR OVERLOADING
Door not open or close due to an obstacle (only for automatic door).
E14
DOOR SWITCHES.
Door not lock when closing is required or not unlock when opening is requi-
red (only for manual door).
E15
MANUAL INTERRUP.
Emergency button pressure where present (only for automatic door).
E16
NO WASHING
Water has not dropped below the tank level probes at the end of the wa-
shing phases.
E17
S
OFTENER SUPPLY
The liquid tank is empty, the filter is dirty, or there is a flow meter failure, in
relation to the liquid softener.
E18
DETERGENT SUPPLY
The liquid tank is empty, the filter is dirty, or there is a flow meter failure, in
relation to the liquid detergent.
E19
S
OLVENT SUPPLY
The liquid tank is empty, the filter is dirty, or there is a flow meter failure, in
relation to the liquid solvent.
E20
NO DISINFECTION
Heating element failure, steam leaks.
E21
STEAMGEN TIMEOUT
Mains tap is closed or the filter is dirty.
E22
SOL. VALVES OFF
No solenoid valve enabled for washing (board setting error).
E30
SOFTENER EMPTY
Softener liquid tank empty - nozzle probe failure.
E31
TEMP NOT REACHED
Heating element failure, steam leaks.
E32
S
T.G. HEAT: TOUT
Heating element failure in the steam generator after a fixed time-out of 20
minutes.
E34
HOT WATER TEMP
Washing hot water temperature is lower than the set one.
E40
STGEN OVERHEATED
Overheating of the steam generator.
E41
LEAK PERIST.PUMP
Liquid leak.
E42
C
LOGGED SIPHON
Drain pipe clogged.
E43
DOOR OPEN
Door opening during a program, due to malfunction of micro-switches.
E44
CALL SERVICE
Maintenance required.
E82
NTC2 PROBE BREAK
Steam generator NTC probe broken or disconnected.
E83
NTC2 PROBE SHORT C
Steam generator NTC probe short-circuited.
E84
C
YCLE "n" INTERRUPTED
Interrupted cycle due to alarm or black-out.
E85
C
YCLE ABORTED
After pressing the reset button while alarm E84 is displayed.
E86
MAX TEMPERATURE
Chamber heating elements failure or chamber PT1000 / PT100 probe
malfunction.
Summary of Contents for DVS
Page 40: ...Operative Instructions 40 Notes ...
Page 41: ...Operative Instructions 41 Notes ...