OMICROM FT
SERVICE MANUAL
21 September 1998
2.20
again the optimum scaling.
2.4.6. Respiration messages.
During the normal operation of Omicrom FT monitor, it can show several alarm
messages regarding the respiration signal (P000PLAC20 board). These
messages are:
1) INTERNAL FAILURE
The self calibration has failed or it is not possible to
communicate with the respiration board.
2) CABLE OFF
The ECG cable is disconnected from the respiration
board.
3) LEAD OFF
The LL lead or RA lead or both leads are
disconnected.
4) APNEA ALARM
No breath has been detected during the apnea
alarm period selected.
2.4.7. System messages.
These are an special type of alarm messages that can produce a total system bad
performance, so when it is detected the normal operation is suspended and a
alarm screen is shown to indicate the type of failure. In this case it is necessary to
turn off the monitor and power on again. If the failure is repetitive contact RGB
MEDICAL DEVICES.
1) INTERNAL FAILURE F01
Illegal software opcode loaded at the display board
microprocessor (P000PLAC14 board).
2) INTERNAL FAILURE F02
Bad RAM addressing at the digital section of the
display board (P000PLAC14 board).
3) INTERNAL FAILURE F03
Memory error at the controtemp board
(P000PLAC04 board).
4) INTERNAL FAILURE F04
Illegal software opcode loaded in the controtemp
board microprocessor (P000PLAC04 board).
5) INTERNAL FAILURE F05
Bad RAM addressing at the digital section of the
controtemp board (P000PLAC04 board).
Summary of Contents for Omicrom FT
Page 3: ......
Page 17: ...OMICROM FT SERVICE MANUAL 21 September 1998 2 10 ...
Page 37: ...OMICROM FT SERVICE MANUAL 21 September 1998 3 10 ...
Page 54: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 6 5 2 2 P000PLAC01 Part list ...
Page 57: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 9 5 2 3 P000PLAC04 Part list ...
Page 59: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 11 5 2 4 P000PLAC05 Part list ...
Page 63: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 15 5 2 6 P000PLAC10 Part list ...
Page 66: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 18 5 2 7 P000PLAC14 Part list ...
Page 69: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 21 5 2 9 P000PLAC19 Part list ...
Page 72: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 24 5 2 10 P000PLAC20 Part list ...
Page 76: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 28 5 3 1 P000PLAC01 Schematics ...
Page 82: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 34 5 3 2 P000PLAC04 Schematics ...
Page 85: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 37 5 3 3 P000PLAC05 Schematics ...
Page 87: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 39 5 3 4 P000PLAC07 Schematics ...
Page 90: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 42 5 3 5 P000PLAC10 Schematics ...
Page 95: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 47 5 3 6 P000PLAC14 Schematics ...
Page 99: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 51 5 3 7 P000PLAC15 Schematics ...
Page 101: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 53 5 3 8 P000PLAC19 Schematics ...
Page 108: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 60 5 3 9 P000PLAC20 Schematics ...
Page 116: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 68 ...