OMICROM FT
SERVICE MANUAL
21 September 1998
2.15
Non invasive pressure signal verification.
Activate the non invasive pressure signal at the Omicrom FT (for some models, it
is necessary to disactivate previously the invasive pressure signal) and activate
the displaying of the oscillating pressure at the cuff.
Inflation pressure.
a) Using a T connector, connect the Omicrom FT, the pressure manometer
and a cuff, that must be wrapped around a rigid cylinder which is
approximately the same size as an arm.
b) Initiate the pressure measurement at the Omicrom FT.
c) Check that the inflation pressure is approximately 160 mmHg for adults
and 100 mmHg for neonates.
Small air leakage verification.
a) Using a T connector, connect the Omicrom FT, the pressure manometer
and a cuff, that must be wrapped around a rigid cylinder which is
approximately the same size as an arm.
b) Initiate the pressure measurement at the Omicrom FT.
c) Do not stop the measurement procedure.
d) Check that the waveform monitored is flat during all pressure steps.
e) Check that the pressure value at each step is accurate.
f) Finally, check that the pressure is below 10 mmHg and that the Omicrom
FT generates a weak signal alarm, so there are not air leakages.
Overpressure cut-out.
a) Using a T connector, connect the Omicrom FT, the pressure manometer
and a cuff, that must be wrapped around a rigid cylinder which is
approximately the same size as an arm.
b) Initiate the pressure measurement at the Omicrom FT.
c) When the cuff pressure is maximum, you have to produce manually an
overpressure higher than 300 mmHg. Check that the safety system that
protect against overpressure is operating and the pressure is limited.
Pressure measurement.
a) Place the cuff correctly around the upper arm and initiate a normal
measurement.
b) Check that the cuff inflation is correct.
c) When the measurement finishes, check that the cuff is deflated
completely and the pressure values are showed at Omicrom FT display.
d) Check that the pressure values measured can be considered as correct
and that consecutive measurements are repeatable.
Non invasive pressure alarms:
a) With no cuff connected initiate a measurement and check that Omicrom
FT shows an alarm message indicating “Inflation timeout” after 45
seconds (20 for neonates).
b) With the cuff wrapped around a rigid cylinder initiate a measurement and
check that Omicrom FT shows an alarm message indicating “Weak
signal”.
Summary of Contents for Omicrom FT
Page 3: ......
Page 17: ...OMICROM FT SERVICE MANUAL 21 September 1998 2 10 ...
Page 37: ...OMICROM FT SERVICE MANUAL 21 September 1998 3 10 ...
Page 54: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 6 5 2 2 P000PLAC01 Part list ...
Page 57: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 9 5 2 3 P000PLAC04 Part list ...
Page 59: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 11 5 2 4 P000PLAC05 Part list ...
Page 63: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 15 5 2 6 P000PLAC10 Part list ...
Page 66: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 18 5 2 7 P000PLAC14 Part list ...
Page 69: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 21 5 2 9 P000PLAC19 Part list ...
Page 72: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 24 5 2 10 P000PLAC20 Part list ...
Page 76: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 28 5 3 1 P000PLAC01 Schematics ...
Page 82: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 34 5 3 2 P000PLAC04 Schematics ...
Page 85: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 37 5 3 3 P000PLAC05 Schematics ...
Page 87: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 39 5 3 4 P000PLAC07 Schematics ...
Page 90: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 42 5 3 5 P000PLAC10 Schematics ...
Page 95: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 47 5 3 6 P000PLAC14 Schematics ...
Page 99: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 51 5 3 7 P000PLAC15 Schematics ...
Page 101: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 53 5 3 8 P000PLAC19 Schematics ...
Page 108: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 60 5 3 9 P000PLAC20 Schematics ...
Page 116: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 68 ...