OMICROM FT
SERVICE MANUAL
21 September 1998
2.11
Electrocardiogram signal verification.
Activate at the monitor the ECG signal and select a scale of 2 mV. Connect the
ECG simulator (Lionheart 3) to the Omicrom FT using the standard cable.
Signal filter checking:
a) Select at simulator a 2 Hz square waveform with an amplitude of 1 mV
and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
b) Select at simulator a 4 Hz sinusoidal waveform with an amplitude of 1
mV and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
c) Select at simulator a 10 Hz sinusoidal waveform with an amplitude of 1
mV and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
d) Select at simulator a 0.5 Hz sinusoidal waveform. Select at the Omicrom
FT the Diagnostic ECG mode and check that at monitor display the peak
amplitude is 50% of monitoring area. Select at the Omicrom FT the
Monitor or OR ECG mode and check that at monitor display the peak
amplitude is 25% of monitoring area.
e) Select at simulator a 50 Hz sinusoidal waveform (or 60 Hz depending on
AC mains supply frequency) with an amplitude of 1 mV and check that at
monitor display a flat line appears (all ECG modes).
ECG signal verification:
a) Select at simulator an ECG waveform with and amplitude of 1 mV and a
frequency of 80 bpm and check that at monitor display the QRS
amplitude is 50% of monitoring area (all ECG modes).
QRS detection:
a) Select at simulator QRS detection signal of 0,5 mV, 70 ms and 80 bpm
and check that the heart rate showed at monitor display is correct (all
ECG modes).
b) Select at simulator QRS detection signal of 0,5 mV, 40 ms and 80 bpm.
Select at Omicrom FT as Patient type neonates and check that the heart
rate showed at monitor display is correct (all ECG modes).
c) Select at simulator QRS detection signal of 0,15 mV, 100 ms and 80
bpm and check that monitor does not detect any QRS (all ECG modes).
Signal gain:
a) Select at simulator an ECG waveform of 1mV and 80 bpm.
b) Select at monitor an ECG scale of 4 mV.
c) Check that the amplitude of the waveform at the monitor display is 25%
of monitoring area.
d) Select at the monitor an ECG scale of 1 mV.
e) Check that the amplitude of the waveform at the monitor display is 100%
of monitoring area.
Lead selection.
a) Select at simulator an ECG waveform of 1mV and 80 bpm.
Summary of Contents for Omicrom FT
Page 3: ......
Page 17: ...OMICROM FT SERVICE MANUAL 21 September 1998 2 10 ...
Page 37: ...OMICROM FT SERVICE MANUAL 21 September 1998 3 10 ...
Page 54: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 6 5 2 2 P000PLAC01 Part list ...
Page 57: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 9 5 2 3 P000PLAC04 Part list ...
Page 59: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 11 5 2 4 P000PLAC05 Part list ...
Page 63: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 15 5 2 6 P000PLAC10 Part list ...
Page 66: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 18 5 2 7 P000PLAC14 Part list ...
Page 69: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 21 5 2 9 P000PLAC19 Part list ...
Page 72: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 24 5 2 10 P000PLAC20 Part list ...
Page 76: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 28 5 3 1 P000PLAC01 Schematics ...
Page 82: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 34 5 3 2 P000PLAC04 Schematics ...
Page 85: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 37 5 3 3 P000PLAC05 Schematics ...
Page 87: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 39 5 3 4 P000PLAC07 Schematics ...
Page 90: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 42 5 3 5 P000PLAC10 Schematics ...
Page 95: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 47 5 3 6 P000PLAC14 Schematics ...
Page 99: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 51 5 3 7 P000PLAC15 Schematics ...
Page 101: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 53 5 3 8 P000PLAC19 Schematics ...
Page 108: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 60 5 3 9 P000PLAC20 Schematics ...
Page 116: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 68 ...